Getting Started
If the vaccine cabinet was shipped to you, please wait at least three hours prior to turning it on.
This will allow enough time for the oil in the compressor to settle after shipping.
Medisafe Plus refrigerators are powered by electric current, which can cause serious injury.
Any installation, cleaning, or troubleshooting must be conducted with the electrical power source
disconnected from the unit.
The following precautions should be taken while the unit is in operation:
• Ensure the appropriate power is supplied to the unit. Medisafe Plus units are designed to operate
at 220- 240V, 50Hz.
• Do not plug several appliances into the same multiple receptacles.
• Electrical components should not come in contact with water.
• Ensure the power cord is not damaged. If the cord becomes damaged, it must be replaced by the manufacturer,
a certied service agent or qualied service professional.
It is strongly recommended that any servicing of the unit be performed by Avem Quirks.
Contact us on 1800 658 120 or servicedesk@avemquirks.com.au
Installation
• Place the Medisafe Plus refrigerator in a location with a at level surface, with reasonable clearance from any
walls and without exposure to direct sunlight.
• Allow clearance to the right, left, back and top. This will reduce power consumption and allow the internal cooling
system to function correctly. The unit is not to be placed in conned areas where air ow and heat dissipation are
restricted.
• Ensure that the appropriate power is supplied to the unit. Medisafe Plus refrigerators are designed to operate at
220-240V, 50Hz. National and local codes provide specic electrical and mechanical installation requirements and
recommendations covering equipment and equipment wiring.
• Plug in and turn on the power to the Medisafe Plus refrigerator. Ensure the power point is secure, that the
refrigerator cannot be accidentally unplugged or switched off.
• Plug in and turn on the power to the AQ Box Cloud Temperature Monitoring system. This is to be plugged into the
same power point as the Medisafe Plus refrigerator. Ensure the plug is secure and cannot be accidentally
unplugged.
When the power is rst turned on, the Min-Max temperature display will ash a “PF” alert as shown below.
This informs the operator there was a loss of power and can be acknowledged by pressing the red “SET”
button on the Min-Max display.
Press the red ‘SET’ button to
mute and stop the alarm.