AGFA XD 10 Series Manual de usuario

DR Detector Calibration
Key User Manual
0134I EN 20210924 1204

Contents
Legal Notice ...........................................................................3
Introduction to this Manual ................................................... 4
Scope .........................................................................5
Disclaimer ................................................................. 6
Calibrating the DR Detector ................................................... 7
About Gain Calibration .............................................. 8
Calibration Frequency ................................................9
Calibration Workflow ...............................................11
Required material ........................................ 12
Preconditions ...............................................13
Procedure .................................................... 14
XD 10, XD⁺10, XD 14, XD⁺14, XD 17 and XD⁺17
Exposure Sequence ...................................... 17
DR 10s and DR 14s Exposure Sequence ........ 20
DR 10e C, DR 14e C and DR 17e C Exposure
Sequence ..................................................... 22
DR 18M and DR 24M Exposure Sequence .... 24
DX-D 10C, DX-D 10G, DX-D 20C, DX-D 20G
Exposure Sequence ......................................25
DX-D 25C Exposure Sequence ...................... 27
DX-D 30C and DX-D 35C Exposure Sequence ...
29
DX-D 40C and DX-D 40G Exposure Sequence ...
31
DX-D 45C, DX-D 45G, DX-D 60C and DX-D 60G
Exposure Sequence ......................................33
Fixed DR Detector (4343R, Pixium RAD 4343)
Exposure Sequence ......................................35
Fixed DR Detector (Pixium RF4343 FL) Exposure
Sequence ..................................................... 37
Problem solving ...................................................................39
DR Calibration Tool Stalls ........................................ 40
Generator error ........................................................41
Generator controls are disabled ............................... 42
ii | DR Detector Calibration | Contents
0134I EN 20210924 1204

Legal Notice
0413
Agfa NV, Septestraat 27, B-2640 Mortsel - Belgium
For more information on Agfa products, please visit www.agfa.com.
Agfa and the Agfa rhombus are trademarks of Agfa-Gevaert N.V., Belgium or
its affiliates. DX-D is a trademark of Agfa NV, Belgium or one of its affiliates.
All other trademarks are held by their respective owners and are used in an
editorial fashion with no intention of infringement.
Agfa NV makes no warranties or representation, expressed or implied, with
respect to the accuracy, completeness or usefulness of the information
contained in this document and specifically disclaims warranties of suitability
for any particular purpose. Products and services may not be available for
your local area. Please contact your local sales representative for availability
information. Agfa NV diligently strives to provide as accurate information as
possible, but shall not be responsible for any typographical error. Agfa NV
shall under no circumstances be liable for any damage arising from the use or
inability to use any information, apparatus, method or process disclosed in
this document. Agfa NV reserves the right to make changes to this document
without prior notice. The original version of this document is in English.
Copyright 2021 Agfa NV
All rights reserved.
Published by Agfa NV
B-2640 Mortsel - Belgium.
No part of this document may be reproduced, copied, adapted or transmitted
in any form or by any means without the written permission of Agfa NV
DR Detector Calibration | Legal Notice | 3
0134I EN 20210924 1204

Scope
This manual contains information for the calibration and maintenance of the
following DR Detectors:
• DR 10e C
• DR 10s
• DR 14e C
• DR 14e G
• DR 14s
• DR 17e C
• DR 17e G
• DR 18M
• DR 24M
• DX-D 10C
• DX-D 10G
• DX-D 20C
• DX-D 20G
• DX-D 25C
• DX-D 30C
• DX-D 35C
• DX-D 40C
• DX-D 40G
• DX-D 45C
• DX-D 45G
• DX-D 60C
• DX-D 60G
• Fixed DR Detector (4343R, Pixium RAD 4343)
• Fixed DR Detector (Pixium RF4343 FL)
• XD 10
• XD⁺10
• XD 14
• XD⁺14
• XD 17
• XD⁺17
DR Detector Calibration | Introduction to this Manual | 5
0134I EN 20210924 1204

Disclaimer
Agfa assumes no liability for use of this document if any unauthorized
changes to the content or format have been made.
Every care has been taken to ensure the accuracy of the information in this
document. However, Agfa assumes no responsibility or liability for errors,
inaccuracies or omissions that may appear in this document. To improve
reliability, function or design Agfa reserves the right to change the product
without further notice. This manual is provided without warranty of any kind,
implied or expressed, including, but not limited to, the implied warranties of
merchantability and fitness for a particular purpose.
Note: In the United States, Federal law restricts this device on order
of a physician for prescription use only.
6 | DR Detector Calibration | Introduction to this Manual
0134I EN 20210924 1204

About Gain Calibration
Regular calibration of the DR Detector assures constant image quality. This
calibration eliminates the effect of sensitivity deviations over the panel and/or
corrects non-unifomity in the X-ray beam. The sensitivity deviations of pixels
across the DR Detector are affected by the quantity and pattern of the effective
X-ray dose on the detector. Depending on the the panel type it will also correct
dead pixels in the acquired images.
8 | DR Detector Calibration | Calibrating the DR Detector
0134I EN 20210924 1204

Calibration Frequency
The DR Detector must be recalibrated at regular intervals. Messages will be
displayed on the NX workstation when recalibration is required.
Perform calibration when exposure conditions have changed significantly and
after a major change to the hardware or software configuration of the X-ray
system.
A DR detector used for veterinary applications has a calibration interval of 12
months, regardless of the model.
Table 1: Calibration interval
DX-D 10C, DX-D 10G, DX-D 20C, DX-D 20G 3 months
DR 18M, DR 24M
DX-D 40C, DX-D 40G, DX-D 45C, DX-D 45G,
DX-D 60C, DX-D 60G
XD 10, XD⁺10, XD 14, XD⁺14, XD 17, XD⁺17
6 months
DR 10e C, DR 14e C, DR 17e C
DR 10s, DR 14s
DX-D 25C, DX-D 30C, DX-D 35C
Fixed DR Detector (Pixium RF4343 FL)
12 months
Fixed DR Detector (4343R, Pixium RAD 4343) 3 months or 12 months,
depending on the DR De-
tector model.
A timely reminder is dis-
played on the NX worksta-
tion.
DR 14e G, DR 17e G no calibration required
Table 2: Using the DR Detector on multiple NX workstations
DR 10s, DR 14s
DX-D 10C, DX-D 10G, DX-D
20C, DX-D 20G
DX-D 25C
DX-D 30C, DX-D 35C
Each NX workstation has its own set of cali-
bration data. The regular calibration must be
performed on each of the workstations.
DR Detector Calibration | Calibrating the DR Detector | 9
0134I EN 20210924 1204

DX-D 40C, DX-D 40G, DX-D
45C, DX-D 45G, DX-D 60C,
DX-D 60G
XD 10, XD 10+, XD 14, XD
14+, XD 17, XD 17+
DR 10e C, DR 14e C, DR 17e
C
A single set of calibration data is used for all
NX workstations on which the DR Detector is
used. Choose one NX workstation that will be
used for calibration and perform the regular
calibration each time on the same NX work-
station.
Do not calibrate the DR Detector on another
NX workstation.
DR 18M, DR 24M
Fixed DR Detector
The DR Detector can only be used on a single
NX workstation.
Note: If the system has multiple DR Detectors, make a note of the ID
that is mentioned in the message on the NX workstation indicating
which of the DR Detectors must be calibrated. The ID is required to
identify the right detector during calibration.
10 | DR Detector Calibration | Calibrating the DR Detector
0134I EN 20210924 1204