4
Information on safety
EN
Unitop
2.2 Intended use
Unitop 3000 part no. 28000
This product may only be used for measuring the level of the following media:
• Fuel oil EL as per DIN 51603-1
• Diesel fuel as per EN 590
• Fatty acid methyl ester (FAME) as fuel oil as per EN 14213
• Fatty acid methyl ester (FAME) as biodiesel as per EN 14214
• Flammable liquids of danger class A III and non-flammable liquids with
the following prerequisites:
- The vapours of the liquids do not attack plastic materials (PA, PS, PE),
Cu, Zn and Sn alloys and elastomers.
- The liquid does not belong to danger classes AI, AII or B.
- Cinematic viscosity < 300 mm²/s.
Unitop AdBlue®part no 28042
This product may only be used for measuring the level of AdBlue®as per DIN
70070 with a specific weight (density) = 1090 kg/m³. The term AdBlue® des-
ignates the same as "NOx reduction agent AUS 32" and "urea solution 32.5
%").
Any use other than the application explicitly permitted in these operating
instructions is not permitted and causes hazards.
Verify that the product is suitable for the application planned by you prior to
using the product. In doing so, take into account at least the following:
• All directives, standards and safety regulations applicable at the installa-
tion site of the product
• All conditions and data specified for the product
• The conditions of the planned application
In addition, perform a risk assessment in view of the planned application,
according to an approved risk assessment method, and implement the
appropriate safety measures, based on the results of the risk assessment.
Take into account the consequences of installing or integrating the product
into a system or a plant.
When using the product, perform all work and all other activities in conjunc-
tion with the product in compliance with the conditions specified in the oper-
ating instructions and on the nameplate, as well as with all directives, stand-
ards and safety regulations applicable at the installation site of the product