AEMC instruments 405 Manual de usuario

405
POWER CLAMP-ON METER
User Manual
ENGLISH
Shop for Power Metering products online at: 1.888.610.7664
www.PowerMeterStore.com

3
CONTENTS
RECEIVING YOUR SHIPMENT.........................................................................................7
ORDERING INFORMATION..............................................................................................7
1PRESENTATION .......................................................................................................8
1.1 THE ROTARYSWITCH ..................................................................................................9
1.2 THE FUNCTION BUTTONS..........................................................................................10
1.3 THE DISPLAY................................................................................................................11
1.3.1 Display Symbols........................................................................................11
1.3.2 Measurement Capacity Exceeded (OL) .....................................................12
1.4 THE TERMINALS ..........................................................................................................13
2THE BUTTONS........................................................................................................ 14
2.1 BUTTON................................................................................................................14
2.2 (YELLOW)BUTTON(SECOND FUNCTION) ...........................................................15
2.3 BUTTON ..............................................................................................................15
2.4 BUTTON..............................................................................................................16
2.4.1 Normal Mode.............................................................................................16
2.4.2 The MAX/MIN Mode + Activation of the HOLD Mode.................................17
2.4.3 Access to the True Inrush®Mode ( set to )...............................17
2.5 BUTTON ..............................................................................................................18
2.5.1 Normal Mode.............................................................................................18
2.5.2 The Hz Function + Activation of the HOLD Mode.......................................18
2.6 BUTTON............................................................................................................19
3USE..........................................................................................................................20
3.1 INSTALLING THE BATTERIES....................................................................................20
3.2 TURNING THE CLAMP-ONMETER ON .....................................................................20
3.3 TURNING THE CLAMP-ONMETER OFF....................................................................20
3.4 CONFIGURATION.........................................................................................................20
3.4.1 Configuring the Maximum Resistance for Continuity..................................21
3.4.2 Auto Power OFF........................................................................................21
3.4.3 Configuring the Threshold for True InRush®Measurement ........................21
3.4.4 Default Configuration.................................................................................22
3.5 VOLTAGE MEASUREMENT (V) ..................................................................................22
3.6 CONTINUITYTEST ................................................................................................23
3.6.1 Lead Resistance Compensation ................................................................24
3.7 RESISTANCEMEASUREMENTΩ..............................................................................24
3.8 DIODE TEST ...........................................................................................................25
3.9 CURRENT MEASUREMENT(A)..................................................................................25
3.9.1 AC Measurement.......................................................................................25
3.9.2 DC or AC+DC Measurement.....................................................................26
3.10 STARTING CURRENT OR OVERCURRENT (True InRush®)
MEASUREMENT...........................................................................................................27
3.11 POWERMEASUREMENTSW,VA,VAR ANDPF........................................................28
3.11.1 Single-Phase Power Measurement............................................................28
3.11.2 Balanced 3-Phase Power Measurement....................................................29
3.12 PHASEROTATION MODE ...............................................................................30
Shop for Power Metering products online at: 1.888.610.7664
www.PowerMeterStore.com

4
3.13 FREQUENCY MEASUREMENT(HZ) ..........................................................................31
3.13.1 Frequency Measurement (V) .....................................................................31
3.13.2 Frequency Measurement (A) .....................................................................31
3.13.3 Frequency Measurement (W) ....................................................................32
3.14 MEASUREMENT OF THE LEVELOFHARMONICS (THD) AND
THE FREQUENCY OF THE FUNDAMENTAL (NETWORK)......................................33
3.14.1 Measurement of THD and Frequency of the Fundamental (V) ...................33
3.14.2 Measurement of THD and Frequency of the Fundamental (A) ...................33
4SPECIFICATIONS ................................................................................................... 35
4.1 REFERENCE CONDITIONS.........................................................................................35
4.2 SPECIFICATIONS UNDER THE REFERENCE CONDITIONS..................................35
4.2.1 DC Voltage Measurement..........................................................................35
4.2.2 AC Voltage Measurement..........................................................................36
4.2.3 AC+DC Voltage Measurement...................................................................36
4.2.4 DC Current Measurement..........................................................................37
4.2.5 AC Current Measurement..........................................................................37
4.2.6 AC+DC Intensity Measurement..................................................................38
4.2.7 True InRush®Measurement.......................................................................38
4.2.8 Continuity Measurement............................................................................39
4.2.9 Resistance Measurement ..........................................................................39
4.2.10 Diode test..................................................................................................39
4.2.11 Active DC Power Measurements ...............................................................40
4.2.12 Active AC Power Measurements................................................................40
4.2.13 Active AC+DC Power Measurements ........................................................41
4.2.14 Measurement of Apparent AC Power.........................................................42
4.2.15 Measurement of Apparent AC+DC Power..................................................42
4.2.16 Measurement of Reactive AC Power .........................................................42
4.2.17 Measurement of reactive AC+DC power....................................................43
4.2.18 Calculation of the Power Factor.................................................................44
4.2.19 Frequency Measurements .........................................................................44
4.2.20 Specifications in THDr ...............................................................................45
4.2.21 Specifications in THDf................................................................................45
4.2.22 Indication of order of the phases................................................................46
4.3 ENVIRONMENTALCONDITIONS ...............................................................................46
4.4 MECHANICAL SPECIFICATIONS ...............................................................................46
4.5 POWERSUPPLY..........................................................................................................47
4.6 COMPLIANCE WITHINTERNATIONALSTANDARDS..............................................47
4.7 ENVIRONMENTALVARIATIONS................................................................................48
5MAINTENANCE....................................................................................................... 49
5.1 WARNING......................................................................................................................49
5.2 CLEANING.....................................................................................................................49
5.3 REPLACEMENTOF THE BATTERIES........................................................................49
6REPAIR AND CALIBRATION.................................................................................. 50
7TECHNICAL AND SALES ASSISTANCE................................................................ 50
8LIMITED WARRANTY ............................................................................................. 51
9WARRANTY REPAIRS............................................................................................ 51
Shop for Power Metering products online at: 1.888.610.7664
www.PowerMeterStore.com

5
Thank you for purchasing a Model 405 Clamp-on Meter.
For best results from your instrument and for your safety, read the enclosed
operating instructions carefully and comply with the precautions for use. These
products must be only used by qualified and trained users.
Meanings of the symbols used on the device
CAUTION - Risk of Danger! Indicates a WARNING and that the operator
must refer to the user manual for instructions before operating the
instrument in all cases where this symbol is marked.
Risk of electric shock. The voltage at the parts marked with this symbol may
be dangerous.
Refers to a type A current sensor. This symbol signifies that application
around and removal from HAZARDOUS LIVE conductors is permitted.
1.5 V battery
The CE marking indicates compliance with European directives
Double insulation or reinforced insulation
In the European Union, this product is subject to a separate collection
system for recycling electrical and electronic components In accordance
with directive WEEE 2002/96/EC
AC – Alternating current
AC and DC – Alternating and direct current
Ground/Earth
Shop for Power Metering products online at: 1.888.610.7664
www.PowerMeterStore.com

6
PRECAUTIONS FOR USE
This device complies with safety standards IEC-61010-1 and 61010-2-032 for
voltages of 1000V in category IV at an altitude of less than 2000m, indoors, with
a degree of pollution not exceeding 2.
These safety instructions are intended to ensure the safety of persons and
proper operation of the device.
The operator and/or the responsible authority must carefully read and clearly
understand the various precautions to be taken in use.
If this instrument is used other than as specified, the protection it provides
may be compromised, thereby endangering you.
Do not use the instrument in an explosive atmosphere or in the presence of
flammable gases or fumes.
Do not use the instrument on networks of which the voltage or category
exceeds those mentioned.
Do not exceed the rated maximum voltages and currents between terminals
or with respect to earth.
Do not use the instrument if it appears to be damaged, incomplete, or not
properly closed.
Before each use, check the condition of the insulation on the leads, housing,
and accessories. Any element of which the insulation is deteriorated (even
partially) must be set aside for repair or scrapped.
Use leads and accessories rated for voltages and categories at least equal
to those of the instrument. If not, an accessory of a lower category lowers
the category of the combined Clamp + accessory to that of the accessory.
Observe the environmental conditions of use.
Do not modify the instrument and only use factory replacement parts.
Repairs and adjustments must be done by approved qualified personnel.
Replace the batteries as soon as the symbol appears on the display of
the unit. Disconnect all leads before opening the battery compartment cover.
Use personal protective equipment when conditions require.
Keep your hands away from the unused terminals of the instrument.
When handling the test probes, alligator clips, and clamp ammeters, keep
your fingers behind the physical guard.
As a safety measure, and to avoid repeated overloads on the inputs of the
device, configuration operations should only be performed when the device
is disconnected from all dangerous voltages.
Shop for Power Metering products online at: 1.888.610.7664
www.PowerMeterStore.com

7
MEASUREMENT CATEGORIES
Definitions of the measurement categories:
CAT IV: Circuits supplying the low-voltage installation of the building.
Example: power lines, meters, and protection devices.
CAT III: Power supply circuits in the installation of the building.
Example: distribution panel, circuit-breakers, fixed industrial machines or
devices.
CAT II: Circuits directly connected to the low-voltage installation.
Example: power supply to household electrical appliances and portable tools.
RECEIVING YOUR SHIPMENT
Upon receiving your shipment, make sure that the contents are consistent with the
packing list. Notify your distributor of any missing items. If the equipment appears
to be damaged, file a claim immediately with the carrier and notify your distributor
at once, giving a detailed description of any damage. Save the damaged packing
container to substantiate your claim.
ORDERING INFORMATION
Clamp-on Meter Model 405 .......................................................... Cat. #2139.50
Includes set of 2 color-coded silicone insulated test leads, test probes and alligator clips,
soft carrying case, 4x1.5V AA batteries and user manual.
Replacement Parts:
Soft Carrying Case .........................................................................................Cat. #2139.72
Set of 2 Color-coded Silicone Test Leads, Test Probes & Alligator Clips ........Cat. #2152.05
Set of 2 Color-coded Silicone Test Leads (Red/Black) 5’................................Cat. #2152.15
Clip – Safety Alligator - Black .........................................................................Cat. #2140.53
Clip – Safety Alligator - Red............................................................................Cat. #2140.52
Black Test Probe ............................................................................................Cat. #5000.30
Red Test Probe ..............................................................................................Cat. #5000.31
Shop for Power Metering products online at: 1.888.610.7664
www.PowerMeterStore.com

8
1
3
4
5
6
2
7
1 PRESENTATION
The Clamp-on Meter Model 405 is a 6000-count professional electrical measuring
instrument that combines the following functions:
Current measurement
Measurement of InRush current / overcurrent (True InRush®)
Voltage measurement
Frequency measurement
Measures harmonic distortion (THD)
Continuity test with buzzer
Resistance measurement
Diode test
Power measurements (W, VA, var and PF)
Phase order indication
Item
Designation See §
1 Jaws with cente
ring marks
(see connection principles) 3.5 to
3.12
2 Physical Guard -
3 Rotary Function Switch 1.1
4 Function Buttons 2
5 Backlit Display 1.3
6 Input Terminals 1.4
7 Trigger -
Figure 1: Clamp-on Meter Model 405
Shop for Power Metering products online at: 1.888.610.7664
www.PowerMeterStore.com

9
1.1 THE ROTARY SWITCH
The rotary switch has six positions. To access the , , , ,
functions, set the switch to the desired function. The functions are described in
the table below.
Figure 2: The Function Rotary Switch
Item Function See §
1 OFF mode – Turns the clamp-on meter off 3.3
2 AC, DC, AC+DC voltage measurement (V) 3.5
3 Continuity test
Resistance measurement Ω
Diode test
3.6
3.7
3.8
4 AC, DC, AC+DC current measurement (A) 3.9
5
Power measurements (W, var, VA) and calculation of the
power factor (PF) AC, DC, AC+DC 3.11
6 Phase rotation order indicator - 3.12
1
2
3
4
5
6
Shop for Power Metering products online at: 1.888.610.7664
www.PowerMeterStore.com

10
1.2 THE FUNCTION BUTTONS
1 2 3
4 5 6
Figure 3: The Function Buttons
Item Function See §
1 Holds the last value on the display
Zero correction ADC/AAC+DC/WDC/WAC+DC
Lead resistance compensation in the continuity and
ohmmeter functions
2.1
3.9.2
3.6.1
2 Selects the type of measurement and configuration
functions (AC, DC, AC+DC)
Selection of single-phase or 3-phase measurement 2.2
3 Enables/disables display backlighting 2.3
4 Enables/disables the MAX/MIN/PEAK mode
Enables/disables the True InRush®mode in current
measurement (A) 2.4
5 Performs Frequency measurements (Hz) and selects
measurement of harmonic distortion (THD)
Display of the powers W, VA, var and PF 2.5
6 Activation of ΔREL mode – Displays differential and
relative values 2.6
Shop for Power Metering products online at: 1.888.610.7664
www.PowerMeterStore.com

11
1.3 THE DISPLAY
5
Figure 4: The Display
Item Function See §
1 Mode selection display 2
2 Display of the measurement value and unit 3.5 to 3.12
3 Display of the MAX/MIN/PEAK modes 2.4
4 Type of measurement (AC or DC) 2.2
5 Total 3-phase power measurements 3.11.2
6 Selected resistance mode display 1.1
7 Low battery indication 5.3
1.3.1 Display Symbols
Symbol Designation
AC Alternating current or voltage
DC Direct current or voltage
AC+DC
Alternating and direct current or voltage
∆
REL Relative value, with respect to a reference
∆
Ref Reference value
Storage of the values and display hold
Max Maximum DC or RMS value
Min Minimum DC or RMS value
PEAK+
Maximum peak value
1
2
3
4
7 6
Shop for Power Metering products online at: 1.888.610.7664
www.PowerMeterStore.com