3Disc FireCR+ Manual de usuario

Service Manual
FireCR+ Computed Radiography Scanner
Doc No.: TM-812
Rev: Aug 2013
Part No.: CR-FPM-01-003
3DISC, FireCR, Quantor and the 3D Cube are trademarks of 3D Imaging & Simulations Corp, South
Korea, and its affiliates. All other trademarks are held by their respective owners and are used in an
editorial fashion with no intention of infringement. The data in this publication are for illustration purposes
only and do not necessarily represent standards or specifications, which must be met by 3D Imaging &
Simulations Corp. All information contained herein is intended for guidance purposes only, and
characteristics of the products and services described in this publication can be changed at any time
without notice. Products and services may not be available in your local area. Please contact your local
sales representative for availability information. 3D Imaging & Simulations Corp. strives to provide as
accurate information as possible, but shall not be responsible for any typographical error.
© Copyright 2013 3D Imaging & Simulations Corp, all rights reserved, printed, and published in South
Korea by 3D Imaging & Simulations Corp.

815, Tamnip-Dong, Yuseong-Gu,
Daejeon, Korea
Tel : 82-42-931-2100
Fax : 82-42-931-2299
Website : www.3DISCimaging.com
E-mail : info@3DISCimaging.com
3DISC Americas
22560 Glenn Dr, Suite 116
Sterling, VA 20164 USA
Tel : 1-703-430-6080
E-mail : info@3DISCimaging.com
3DISC Europe
Gydevang, 39-41, 3450 Alleroed, Denmark
Tel : 45-88-276-650
E-mail : info@3DISCimaging.com

Service Manual FireCR+Computed Radiography Scanner
FireCR+| Contents 3
Contents
Components .......................................................................................................4
Installation .........................................................................................................5
Table Top................................................................................................................................................... 5
Wall Mount ............................................................................................................................................... 6
Connections .............................................................................................................................................. 8
Cassette and IP Visual Inspection ............................................................................................................. 8
Calibration ..........................................................................................................9
Calibration Geometry................................................................................................................................ 9
Calibration Procedure ............................................................................................................................. 10
Reader Diagnostics ...........................................................................................12
Scanner Control.......................................................................................................................................12
Diagnostics Window................................................................................................................................13
Parts Removal/Replacement ............................................................................14
Top Cover ................................................................................................................................................14
Internal Cover .........................................................................................................................................16
Main Board..............................................................................................................................................17
Core Board .............................................................................................................................................. 18
BLDC Board.............................................................................................................................................. 19
Cassette Lock Motor Board..................................................................................................................... 21
Optic Plate...............................................................................................................................................22
Aligner Motor..........................................................................................................................................23
Y-Axis Motor............................................................................................................................................25
Fiber Bundle ............................................................................................................................................27
Internal Frame......................................................................................................................................... 30
RFID Board............................................................................................................................................... 32
Power Board............................................................................................................................................ 33
Troubleshooting ...............................................................................................35
Image Quality .......................................................................................................................................... 35
Mechanical Noise....................................................................................................................................39

Service Manual FireCR+Computed Radiography Scanner
4 Components | FireCR+
Components
Recommended Computer Requirement
Operation System
Microsoft Windows 7
CPU
Core Duo / Core2 Processor
Memory
RAM 4GB or more
Hard Disk
300GB Free Hard Disk Space
Network
1Gbps Ethernet
Video
32 bit Color Display
Video Resolution
1280 x 1024
Item
Qty
FireCR+ Computed Radiography Scanner
1
Fire Kit (Optional Product Identification Card)
1
14” x 17” Cassette
1
10” x 12” Cassette
1
14” x 17” Imaging Plate (Installed in the cassette)
1
10” x 12” Imaging Plate (Installed in the cassette)
1
IP Extractor
1
USB 2.0 Interface Cable
1
Power Adapter
1
Power Cord
1
Transportation Case
1
FireCR+ User Manual
1

Service Manual FireCR+Computed Radiography Scanner
FireCR+| Installation 5
Installation
Table Top
1. Install on a flat and stable surface.
2. Adjust the feet underneath the scanner to ensure stability.
3. Allow 50cm (20 inches) of free space at the front for cassette insertion and 20cm (8 inches) in
the back for cables and power switch access.

Service Manual FireCR+Computed Radiography Scanner
6 Installation | FireCR+
Wall Mount
1. Turn reader on its side. Note: Do not turn upside down!
2. Remove adjustable feet.
3. Attach brackets to the scanner. Note: Washers go between the brackets and the bolts.

Service Manual FireCR+Computed Radiography Scanner
FireCR+| Installation 7
4. Secure wall mount bracket to wall. Note: Make sure that the wall bracket is mounted level.
5. Carefully attach the scanner to the wall bracket. Note: Make sure all four bracket hooks are
securely attached to the wall mount bracket.

Service Manual FireCR+Computed Radiography Scanner
8 Installation | FireCR+
Connections
1. Connect power supply and USB cable.
Cassette and IP Visual Inspection
1. Remove the IPs from the cassettes by hooking and pulling the plate using the IP Extractor.
2. Visually inspect for defects (air bubbles, scratches, dust, etc.)
3. Inspect cassette housing for physical damage.

Service Manual FireCR+Computed Radiography Scanner
FireCR+| Calibration 9
Calibration
Calibration Geometry
Note: Exposure must always cover
the entire cassette.

Service Manual FireCR+Computed Radiography Scanner
10 Calibration | FireCR+
Calibration Procedure
This screen will appear:
Step 1: Auto alignment
Step 2: Erase
Run an erase cycle to remove any residual radiation that may be left on the phosphor.
Step 3: Scan Blank
Without exposing the cassette, insert it into the reader and press the “Scan Blank” button. Note:
Use the same cassette for the entire calibration.
Step 4, 5, & 6: Scan Low, Medium, & High Dose
Expose the entire cassette at the recommended values and if the value is out of range adjust mAs
until the numbers are green. Note: mAs and calibration values are directly proportional, so a
5% increase in mAs will equal a 5% increase in calibration value, etc.
Step 7: Calibration!
Press and wait until the software confirms that the calibration was successful. Note: Cancelling
the calibration before completion will force you to start over.
Tabla de contenidos